HART Procedures Procedural Considerations
Release 0 34-ST-25-20 MC Toolkit User Manual 67
7/03
Calibration - Apply Values (non-Honeywell Transmitter)
Select the OK button. The
display at right appears.
The Current Applied
Process Value field shows
the value of the Process
Value. A new sample of the
input level is displayed each
time the user selects the
Read New Value button.
Adjust the process variable to
the desired LRV while
repeatedly selecting the
Read New Value button to
monitor and verify the input
value.
When the Process Variable is
stabilized at the desired input
level, select the Set New
Value button.
The popup at right appears.
Select the OK button to write
the input value as the LRV
calibration value; the popup
message box will disappear.
Apply
Values
(20 mA-
URV))
Select the 20 mA button. The
popup at right appears.
Adjust the process variable to
the desired LRV while
repeatedly selecting the
Read New Value button to
monitor and verify the input
value.
When the Process Variable is
stabilized at the desired input
level, select the Set New
Value button.
This popup message appears.
Select the OK button to write
the input value as the URV
calibration value.
Vedere la pagina 74