HART Procedures Procedural Considerations
Release 0 34-ST-25-20 MC Toolkit User Manual 49
7/03
Honeywell HART Diagnostics/Service Menu Procedures
Device
Status
The Critical status block lists
the conditions that render the
transmitter device inoperable.
The Non-Critical status
block lists advisories of
conditions that are
noteworthy, but that do not
cause non-operability.
Refer to the section on
Messages and Diagnostic
Codes.
Write
Protect
The Write Protect function
is available for the
Honeywell STT 250
Temperature Transmitter.
Provides write protection for
all parameters.
To enable write protection,
select
Write protected, then
select the
Send button.
To disable Write Protection,
select Not write protected,
enter the password, and then
select the Send button.
The password can be
changed only when Not
Write Protected.
To Change Password,
select the Start button, type
in the new password, and
then select the
Send button.